Eppendorf D30 BioPhotometer User Manual
Page 57

57
Maintenance
Eppendorf BioPhotometer
®
D30
English (EN)
Have the filter recertified by Eppendorf Service after 2 years.
8.3.1.2 Checking the wavelength systematic error
To check the wavelength systematic error, proceed as follows: Measure the 3 test filters at the
corresponding wavelength.
4. Result display after measuring a test filter for
checking the photometric accuracy.
Measure the other two test filters A2 and A3.
5. Results display after measuring all 3 test filters
for testing the photometric accuracy.
With the and keys, you can view again the
results of the different test filters. Press [Finish]
to complete the test.
6. Compare the average values and variation
coefficients with the supplied table.
If the measured values do not agree with the
permitted range of values, contact Eppendorf
Service.