2 mastercycler, 3 thermomixer and thermostat – Eppendorf Temperature Verification System User Manual
Page 35
Advertising

Temperature Verification System USB – Single-channel — Operating manual
35
9.3.2
Mastercycler
9.3.3
Thermomixer and ThermoStat
RT = room temperature
9.3.4
ThermoMixer C, ThermoMixer F1.5, ThermoMixer FP and ThermoStat C
Mastercycler family
Temperature
Error limits
Mastercycler nexus
35 °C
±0,6 K
95 °C
±1,0 K
Mastercycler ep
35 °C
±0,6 K
95 °C
±1,0 K
Mastercycler pro
35 °C
±0,6 K
95 °C
±1,0 K
Temperature
Thermomixer
comfort
Thermomixer
compact
ThermoStat plus
0 °C
———
———
±1,0 K
10 °C below RT
±2,0 K
———
———
37 °C
±0,5 K
±1,0 K
±0,5 K
56 °C
±2,0 K
±2,0 K
±1,0 K
75 °C
±2,0 K
±2,0 K
±1,0 K
90 °C
±2,0 K
±2,0 K
±1,0 K
Temperature
Error limits
20 °C-45 °C
±1,0 K
<20 °C and >45 °C
±1,0 K
Advertising