5 operation, Operation – Eppendorf Research pro User Manual
Page 20

20
Free jet dispensing
Free jet dispensing is possible at volumes greater than
≥
20 µL. The small drop that forms after
each free jet dispensing of a partial volume in the dispensing mode is part of the following
partial volume.
For greater precision and accuracy:
– Activate return stroke (see procedure) in the free jet for free jet dispensing.
– Activate return stroke with the pipette tip against the tube wall for wall dispensing.
First dispensing step is
executed.
Second dispensing step is
executed.
Last dispensing step is
executed.
Here: special acoustic signal
is emitted.
Blank step with special
acoustic signal.
Blow-out is discarded.
No dispensing step.
If the Actuate key is held
down during dispensing, the
device does not return
to
the basic position until the
key is released.
Pipette is once again ready to
aspirate liquid.
Free jet
discharge
Discharge
to wall
Free jet
discharge
Discharge
to wall
Free jet
discharge
Discharge
to wall