7 troubleshooting 7, Troubleshooting – Eppendorf Research pro User Manual

Page 31

Advertising
background image

31

Error

Cause

Solution

Pipette is dripping;
the volume
dispensed is
incorrect.

–

The tip may be loose.

–

There may be foreign
matter between the
pipette and the tip.

–

The piston, nose cone and
cylinder may be damaged
or contaminated.

–

The lower part may be
loose.

–

For multi-channel model:
The O-ring may be
damaged.

–

Attach the tip firmly. Make sure that
an Eppendorf tip is used.

–

Wipe the pipette. Ensure that the tip
is protected from dust.

–

Treat the lower half of the pipette as
described in Sec. 6 and Part B
"Ordering Information/service
parts".

Residual liquid is
in the tip;
dispensing is
incomplete.

–

See above.

–

The dispensing speed
may be too high.

–

The incorrect pipetting
option may have been
selected.

–

See above.

–

The procedure for checking the
function is described in Sec. 5,
"Operation".

Motor stops
during the
dispensing
procedure.

–

The battery may be run
down.

–

The pipette may be
heavily contaminated.

–

Discard the dispensing as it is
incorrect. Recharge the battery
(Sec. 3.3) or treat the pipette as
described in Sec. 6.

Multi-channel:
Pipette doesn’t
aspirate liquid.

(Continued
on next page)

The magnetic coupling on the
multi-channel lower part no
longer works.

In the case of the Research
pro Pipette 4860, strong
impact stress can result in a
separation of the magnetic
connection between the
spindle and the piston
actuator. As a result of this
separation, the pipette no
longer aspirates liquid, despite
movement of the motor.

In order to reestablish the connection,
proceed as follows:

– Separate the ejector from the lower

part.

– To do this, press the lower part

against a surface until it springs in.
Simultaneously press the stop clutch
inward with side "B" of the universal
tool. The ejector cover panel springs
up.

Troubleshooting

7 Troubleshooting

7