3 short mix, 5 temperature control, 1 temperature control without mixing process – Eppendorf ThermoMixer FP User Manual
Page 21: Short, Temperature control 5.5.1, Temperature control without mixing process

Operation
Eppendorf ThermoMixer
®
F0.5/F1.5/F2.0/FP
English (EN)
21
5.4.3
Short
Mix
Use the Short
Mix function for mixing for a short while without temperature control.
1. Set the mixing frequency with the speed arrow keys.
2. Keep the
short key pressed.
The mixing process continues as long as the short key will be pressed.
3. In order to end Short
Mix, release the
short key.
5.5
Temperature control
The Eppendorf ThermoMixer
F0.5/F1.5/F2.0/FP can be used for temperature control in a
range of 4
°C above the ambient temperature to 100
°C.
5.5.1
Temperature control without mixing process
1. To switch off the mixing function, use the speed arrow keys to select the 0
rpm
setting
(▼
below
300
rpm or ▲
above
1
500
rpm or 2
000
rpm).
2. Using the temp arrow keys set the temperature.
• The device immediately starts to perform temperature control.
• When the set temperature is not changed, the actual temperature flashes on the
display and the device does not perform temperature control.
3. To manually start the temperature control procedure, press the start/stop key.
• The symbol flashes on the display.
• The display shows the actual temperature/set temperature.
NOTICE! Damage to electronic components due to condensation.
Condensate can form in the device after it has been moved from a cool
environment to a warmer environment.
After installing the device, wait at least for 3
h. Only then connect the device
to the mains.
• When the actual temperature flashes on the display, the device is not in
temperature control mode operation.
• As soon as the set temperature is changed with the temp arrow keys, the
device begins to perform temperature control.
• When the set temperature has been reached, the display only shows one
value.