Eppendorf ThermoMixer FP User Manual
Page 13

Safety
Eppendorf ThermoMixer
®
F0.5/F1.5/F2.0/FP
English (EN)
13
WARNING! Contamination due to opening seals of consumables.
In the following cases, the seals of tubes can spring open. Sample material can
escape.
• high vapor pressure of the content
• improperly sealed cover
• damaged sealing lip
• improperly fastened foil
Always check that consumables have been sealed tightly before use.
WARNING! Injury from sample material being thrown out.
Sample material can be thrown out of open, improperly sealed or unstable tubes
and plates.
Only mix in closed tubes and closed plates.
Observe the nationally prescribed safety environment when working with
hazardous, toxic and pathogenic samples. Pay particular attention to personal
protective equipment (gloves, clothing, goggles etc.), extraction, and the
safety class of the lab.
CAUTION! Poor safety due to incorrect accessories and spare parts.
The use of accessories and spare parts other than those recommended by
Eppendorf may impair the safety, functioning and precision of the device.
Eppendorf cannot be held liable or accept any liability for damage resulting from
the use of incorrect or non-recommended accessories and spare parts, or from
the improper use of such equipment.
Only use accessories and original spare parts recommended by Eppendorf.
CAUTION! Crush hazard due to moving parts.
Do not replace any consumables during the mixing process.
Put on the ThermoTop or Lid prior to the mixing process.
Do not remove the ThermoTop or Lid during the mixing process.
NOTICE! Caution! Strong vibration.
When mixing at high speeds, items located near the device may be moved by the
vibrations of the work surface and, e.g., fall off the work table.
Do not place easily movable items near the device or secure them adequately.