26
Maintenance
Eppendorf ThermoMixer
®
F0.5/F1.5/F2.0/FP
English (EN)
Auxiliary equipment
• Lint-free cloth
• Mild, soap-based lab cleaner
• Dist. water
Cleaning the Eppendorf ThermoMixer
F0.5/F1.5/F2.0/FP
1. Switch off the Eppendorf ThermoMixer
F0.5/F1.5/F2.0/FP and disconnect it from the
power supply.
2. Clean all of the outer parts of the Eppendorf ThermoMixer
F0.5/F1.5/F2.0/FP with a
mild soap solution and a lint-free cloth.
3. Wipe off the soap solution with dist. water.
4. Dry all cleaned parts.
7.3
Disinfection/Decontamination
Auxiliary equipment
• Lint-free cloth
• Disinfectant
1. Switch the Eppendorf ThermoMixer
F0.5/F1.5/F2.0/FP off and isolate from the power
supply.
2. Allow the device to cool down.
3. Clean the device (see Cleaning on p. 25).
4. Select a disinfection method which complies with the legal requirements and
regulations applicable to your range of application.
5. Wipe the surfaces with the lint-free cloth and disinfectant.
DANGER! Electric shock as a result of penetration of liquid.
Switch off the device and disconnect the power plug before starting cleaning
or disinfection work.
Do not allow any liquids to penetrate the inside of the housing.
Use closed tubes and closed plates.
Do not spray clean/spray disinfect the housing.
Only plug the device back in if it is completely dry, both inside and outside.