7.2 Disinfection/Decontamination

Auxiliary equipment

Lint-free cloth

Disinfectant.

1. Switch the ThermoMixer F1.5/ThermoMixer FP off and isolate from the power supply.
2. Allow the device to cool down.
3. Clean the device (see Cleaning on p. 23).
4. Select a disinfection method which complies with the legal requirements and regulations

applicable to your range of application.

5. Wipe the surfaces with the lint-free cloth and disinfectant.

7.3 Decontamination before shipment

If you are shipping the device to the authorized Technical Service for repairs or to your authorized

dealer for disposal please note the following:

7.4 Verification of temperature control

To verify the temperature accuracy of the thermoblock, use the Eppendorf Temperature

Verification System – Single Channel. In combination with the temperature sensor for the

ThermoMixer F1.5/ThermoMixer FP the exact temperature in the thermoblock can be measured.
Details on the verification process with the Eppendorf Temperature Verification System –Single

Channel can be found in the corresponding operating manual.

DANGER!

DANGER! Electric shock as a result of penetration of liquid.

Switch off the device and disconnect the power plug before starting cleaning or

disinfection work.

Do not allow any liquids to penetrate the inside of the housing.

Use closed tubes and closed plates.

Do not spray clean/spray disinfect the housing.

Only plug the device back in if it is completely dry, both inside and outside.

WARNING!

WARNING! Risk to health from contaminated device

1. Observe the information in the decontamination certificate, You find it as a PDF file

on our website (www.eppendorf.com/decontamination).

2. Decontaminate all the parts you would like to dispatch.
3. Include the fully completed decontamination certificate in the package.