5 introduction, 1 scope of manual, 2 ln2 back-up system – Eppendorf LN2 Backup System User Manual

Page 13: 3 product versions, Introduction 5.1, Scope of manual, Back-up system, Product versions, 5introduction 5.1 scope of manual, 2 ln

Advertising
background image

Introduction

New Brunswick LN2 Back-up System

English (EN)

13

5

Introduction

5.1

Scope of manual

This manual provides the user with the necessary information for the installation and operation of the LN

2

Back-up System for New Brunswick Innova

®

and Premium ranges of Ultra Low Temperature Freezers.

5.2

LN

2

Back-up System

The LN

2

Back-up System system is designed to provide a means of maintaining the contents of a New

Brunswick Ultra Low Temperature Freezer at a preset temperature should the power or the refrigeration
system fail.

The freezer is fitted with a platinum resistance thermometer (referred to as a temperature sensor or probe)
to measure the internal temperature, a solenoid valve and a door switch.

A freestanding control unit and cables are provided. Transfer lines and connectors are supplied to connect
between the solenoid valve and the LN

2

low-pressure storage Dewar. Dewars are not provided.

The backup control unit is plugged into the power supply and a temperature selected between -50

°C

and

-85

°C. When the temperature within the freezer rises above the selected backup temperature, the solenoid

valve switches on and LN

2

is continuously injected into the body of the freezer.

The solenoid valve continues to inject LN

2

until the temperature within the freezer drops below the preset

value selected on the

Temp Set dial. If a power failure occurs, a battery within the back-up unit will

continue to operate the back-up system for up to 48 hours.

When the door or lid of the freezer is opened, the door switch is activated to prevent additional gas
injections while the freezer is open. This prevents the freezer user from being burned by the extremely cold
gas stream.

5.3

Product versions

A low-pressure liquid nitrogen storage Dewar with a pressure-regulated liquid withdrawal
port is required. Liquid nitrogen must be used to ensure proper operation.

Back-up Systems

Freezer type

Part number

Innova 115 V / 220 V 60 Hz CGA Bottle Fitting

U9044-0002

Innova 230 V 50 Hz BS341 Bottle Fitting

U9044-0004

Premium 115 V / 220 V 60 Hz CGA Bottle Fitting

U9044-0006

Premium 230 V 50 Hz BS341 Bottle Fitting

U9044-0008

Advertising