8 ln2 supplies, 9 spare parts, Supplies – Eppendorf LN2 Backup System User Manual

Page 31: Spare parts, 8 ln

Advertising
background image

Technical data

New Brunswick LN2 Back-up System

English (EN)

31

Gas consumption was measured at the indicated temperatures, in the center of the cabinet and in an
ambient temperature of 23 ºC.

9.8

LN

2

Supplies

Consult your local gas supplier for suitable LN

2

supplies. A low-pressure liquid nitrogen storage Dewar (or

other suitable storage tank) with a liquid withdrawal port is required to properly utilize the cooling
properties of the liquid nitrogen. The liquid withdrawal port must have an adjustable valve with a pressure
gauge. The withdrawal pressure should be set to between 6 to 8 psig (0.44 to 0.55 bar).

9.9

Spare parts

Parts should only be replaced/installed by a qualified service engineer.

Consumption in kg/hr
(lbs/hr):

at -60 °C

at -70 °C

at -80 °C

Upright freezers
(empty)

U725 Innova

2.62 (5.76)

2.75 (6.05)

3.0 (6.6)

U535 Innova

2.05 (4.51)

2.16 (4.74)

2.11 (4.63)

U101 Innova

1.13 (2.49)

1.38 (3.03)

1.79 (3.93)

U410 Premium

2.0 (4.4)

2.2 (4.84)

2.6 (5.72)

U570 Premium

2.25 (4.95)

2.7 (5.94)

3.3 (7.26)

Chest freezers (empty)

C760 Innova

2.85 (6.3)

3.0 (6.6)

3.2 (7.0)

C585 Innova

2.6 (5.72)

2.85 (6.3)

3.1 (6.82)

C660 Premium

2.67 (5.86)

2.83 (6.22)

3.03 (6.63)

C340 Premium

1.99 (4.37)

2.07 (4.58)

1.84 (4.04)

Description

Part Number

LN

2

Control Unit

P0625-0570

Power Supply, 18

V,

45

- Watt

P0625-0500

Battery 12

V

7

Ah - Control unit

K0480-0020

Fuse 20

mm

2

A - Control

K0380-0610

PT100 Temperature Sensor Assembly

P0625-1410

Magnetic Door Switch Sensor Assembly

P0625-1420

Door Switch Magnet

K0400-0631

Magnetic Door Switch Adaptor Plates, Innova and
Premium U700 only

P0625-0320

Advertising