S41i CO2 Incubator Shaker — Operating manual
Abb. 12:
Rear view (back panel removed)
Fig. 12:
Rear view (back panel removed)
2. Use the tubing clips provided to eliminate CO
2
leaks.
3. Plug the mains/power cord into its receptacle on the right side of the control box.
4. After verifying that your supply voltage matches the voltage specified for your incubator/
shaker and that the ON/OFF switch is OFF, plug the cord into your mains/power supply outlet.
5. Press the Auto-Zero HEPA filter gently into the white plastic filter socket on the right side of
the control box.
1 ON/OFF switch
4 Inline gas regulators (access from the side)
2 Door handle
5 Access port
3 See Control box (right side) figure for details.
WARNING!
Risk of electrical hazard
Before making electrical connections, verify that your mains/power supply voltage matches
the voltage of your incubator/shaker and that the on/off switch is in the OFF position.
4
Installation