4.7
Install flask clamps
Clamps for 2
–
6
L flasks are shipped with an additional girdle to keep the flasks in place. To install 2
–
6
L
clamps:
WARNING!
Do not overfill flasks (i.e., more than 50% of the flask volume).
Eppendorf flask clamps are used on a variety of shaker platforms. Flat Phillips head screws of
different lengths and thread pitch are used to secure the clamp (see
All clamps are shipped complete with hardware.
Abb. 4-1: Double girdle clamp
Fig. 4-1:
Double girdle clamp
1. Place the clamp on the platform and secure it in
place with the correct type of screws.
2. Place the loose girdle around the upper portion of
the clamp body so that it is held in place by the
clamp legs.
3. Insert the flask into the clamp.
The instructions and illustration above also apply to 2800
ml Fernbach flask clamps.
1 Clamp mounting holes
With screws
3 Lower girdle with girdle tubes
Prevents flask from spining
2 Upper girdle with girdle tubes
Secures flask within the clamp
4 Clamp body (legs and base)
1
2
4
3