3 main illustration, 1 reference 2 single-channel, Main illustration 2.3.1 – Eppendorf Pipette Reference 2 User Manual

Page 9: Reference 2, Single-channel

Advertising
background image

Product description

Eppendorf Reference

®

2

English (EN)

9

2.3

Main illustration

2.3.1

Reference 2

single-channel

1

Volume display
Volume display with four-digit counter for variable pipettes.
Decimal places are marked by a hyphen.

2

Color code
The color code on the control button and the trays of the
matching epT.I.P.S. pipette tips are in the same color.

3

Control button
Sets and dispenses the volume, and ejects the pipette tip.

4

Volume locking button
Locks the control button volume setting ring. Press a volume
locking button at the left or right to set the volume.

5

User adjustment
Upon delivery, the opening is sealed using a gray plastic
adjustment seal (ADJ).

6

User adjustment display
Set to "0" before delivery.

7

Serial number
Serial number for upper and lower parts.

8

Nominal volume

9

Ejector sleeve
Ejects the pipette tip. Press the control button down all the
way.

10

Spring-loaded tip cone

The spring-loading action optimizes the force required to
attach and eject tips (no spring-loaded action for 2

mL

to

10

mL). For 2

mL to 10

mL pipettes, an exchangeable

protection filter in a sleeve is located in the tip cone.

11

Pipette tip

We recommend using epT.I.P.S.