HPLC adapter — Instructions for use

2

You can calculate the g-force (rcf) for a special adapter using the following formula according to

DIN 58 970:
rcf = 1.118 * 10

-5

* n

2

* r

max

, n: rotational speed (rpm), r

max

: max. centrifugation radius (cm)

Example: The max. centrifugation radius of the HPCL adapter is 8.3 cm in the F-45-18-17-Cryo

rotor. A max. g-force of 3,500 x g is reached at 6142 rpm.

Ordering information

Max. allowed tube lengths without cap

33 mm

Adapter bore (Ø)

11.8 mm

Material

Polypropylene

Order No.

(International)

Order No.

(North America)

Description

5427 708.006

5427708006

HPLC adapter

18 pcs.

Your local distributor: www.eppendorf.com/worldwide

Eppendorf AG · 22331 Hamburg · Germany · Tel: +49 40 538 01-0 · Fax: +49 40 538 01-556