Instructions for use
Rotor FA-45-48-11, FA-45-30-11, FA-45-24-11-Kit, FA-45-16-17
English (EN)

3

Operation

3.1

Closing the QuickLock rotor lid

3.2

Retrofitting older centrifuges

You can view the software version on your device’s display shortly after switching on the
device. Retrofitting may only be carried out by a trained service technician. If you require
retrofitting, please contact your dealer.

NOTICE! The rotor lid screw may become loose if it is improperly handled.
The lock on aerosol-tight QuickLock rotor lids may become loose if it is used to hold
the rotor.

Never hold or transport the locked rotor on the rotor lid screw.

Always pick up the rotor using both hands.

1. Check that the outer sealing ring is seated

correctly in the groove.

2. Vertically fit the rotor lid on the rotor.
3. To lock the rotor, turn the red rotor lid

screw clockwise, beyond an audible click
until the stop is reached.

The rotor is not closed properly until an audible click is heard!

If the closing system is difficult to operate apply a thin layer of pivot grease to the
pins in the rotor lid screw.

Rotor

Operation possible without retrofit

Retrofit of older devices possible

FA-45-24-11-Kit
FA-45-30-11

from software version 1.0 onwards

FA-45-48-11
FA-45-16-17

5430:
from software version 4.4 onwards

5430:
from serial number 10 000
onwards

5430 R

• with software version 1.5

• from software version 2.2 onwards

5430 R:
from software version 1.0 onwards