1 loading the rectangular bucket, 2 loading the plate buckets, 4 maintenance – Eppendorf A-4-81-MTP/Flex Rotor User Manual

Page 2: 1 cleaning, 5 technical data

Advertising
background image

Instructions for use
Rotor A-4-81, Rotor A-4-81-MTP/Flex
English (EN)

Your local distributor: www.eppendorf.com/contact

Eppendorf AG · 22331 Hamburg · Germany

[email protected] · www.eppendorf.com

Eppe

ndo

rf

®

and

the Ep

pend

or

f logo

ar

e r

egiste

re

d

tr

ade

m

ar

ks of

Epp

e

ndo

rf A

G.

5810 718

.961-

14/042013

3.2.1

Loading the rectangular bucket

Insert the adapters into the rectangular bucket with the black clamp facing the
"Eppendorf" logo on the bucket.

For tubes with a length of more than 100 mm perform a manual swing-out test with
empty tubes.
The tubes must not come into contact with the rotor cross.

3.2.2

Loading the plate buckets

Observe the maximum load and the maximum height when using stacked plates.

4

Maintenance

4.1

Cleaning

• All buckets, buckets, caps and adapters can be autoclaved (121°C, 20 min).
• The rotor cross is not autoclavable.

Clean the rotor and accessories with a mild cleaning agent.

5

Technical data

Rectangular bucket
500 mL

Bucket for concical
tubes 7 × 50 mL

MTP/Flex bucket

Max. g-force (rcf)

3 220 × g

3 220 × g

2 900 × g

Max. speed (rpm)

4 000 rpm

4 000 rpm

4 000 rpm

Max. load per bucket
(adapter, vessel and
content)

780 g

7 × 75 g

380 g

Max. centrifugation
radius

18 cm

18 cm

16.3 cm

Max. loading height/
vessel length

Please note the data in the operating
manual of the centrifuge.

60 mm

Advertising
This manual is related to the following products: