PCR Tube Strips and Cap Strips — Instructions for use

3

Application

Instructions for use

Example: real-time PCR Tubes and Masterclear Cap Strips.

1. Use gloves.

2. Place the Cap Strips with the optical surface downwards into the tube strip or the 96

PCR plate. To do this, use the alignment with the markings A and H.

3. Press the Cap Strips firmly onto the tubes.

For assistance use the accompanying tools. With the flat side the lids can be

selectively pressed onto the tube (see Fig.).

4. Check the right seating of the Cap Strips on the tube strips or the 96 PCR plates as

usual.

5. Because of the tight seal it is more difficult to open the Masterclear Cap Strips.

Therefore, move the lid strips sideways when pulling off after the PCR run. Thus, lid

after lid can be loosened in the well and taken off afterwards.

4

Technical data

Instructions for use

* Certificates or test procedures and further information on request.

A

A

H

H

Material

Polypropylene, autoclavable (121 °C, 20 min) in open

state.

Eppendorf PCR clean

Certified free of RNase, DNase, human DNA and PCR

inhibitors*.

Centrifugability

Can be centrifuged up to 10,000 x g in the 45° fixed-angle

rotor. 

Test conditions: 4 °C, 30 min, diluted saline solution with a

density of 1.2 g/mL.
PCR Tubes are not suitable for centrifugation of organic 

solvents. In case of doubt please contact Eppendorf.

Storage

Protect from direct sunlight and UV-light.

Store dry at room temperature.

Operational temperature -80 °C to +100 °C, autoclavable at 121 °C.