• Calibration Procedure

(for models equiped with 4-20mA

output option)

R4

R9

ST

I 20

01

R3

R5

J1

R7

R1

R8

D1

R2

C1

C2

D2

C3

Q1

U2

R6

U3

+24vdc

4 to 20ma Out

SPA

N

OF

FS

ET

RE

D

BLK

WH

OU

T

24V

Offset (20 mA)

Adjustment

Zero (4 mA)

Adjustment

1.

Set DC power supply voltage at 24VDC.

2.

Connect ammeter in series with loop.

3.

Move float to the bottom position

4.

Adjust the Offset potentiometer until meter reads 4ma

5.

Move float to top position

6.

Adjust the Span potentiometer until meter reads 20ma.

7.

Repeat steps 3 – 6 for final adjustment.