C.B.S. Scientific SG Series User Manual

Page 20

Advertising
background image

C.B.S. œ Scientific

20

Adjustable Height Sequencer

4.5

Recommended Buffers and Reagents

Pre-mixed acrylamide stock solutions are the method of choice. Use according to
manufacturer’s instructions.

EDTA solution
(0.5 M EDTA, pH 8.0, 100ml)
Na

2

EDTA•2H

2

O (FW 372.2)

0.5M

18.6g

Deionized

H

2

O

70.0ml

Dissolve

10M NaOH. Adjust to pH 8.0.

∼5.0ml

Deionized

H

2

O

to

100.0ml

Store at room temperature


Tris-borate-EDTA buffer, 5X
(5X TBE, pH

∼8.3, 1 liter)

Tris

(FW

121.1)

0.45M

54.5g

Boric acid (FW 61.83)

0.45M 27.8g

EDTA solution (0.5 M, pH 8.0)

0.001M 20.0ml

Deionized

H

2

O

to

1000.0ml

Do not adjust pH. Store at room temperature


Acrylamide stock solution
(40%, 19:1 acrylamide:bisacrylamide, 250ml)
Acrylamide

(FW

71.08)

95.0g

Bis*

(FW

154.2)

5.0g

Deionized

H

2

O

to

250.0ml

Store at 4

°C protected from light.

*N,N’

Methylenebisacrylamide


Ammonium persulfate solution
(10% APS, Initiator, 1ml)
APS

(FW

228.2) 10%

0.1g

Deionized

H

2

O

to 1.0ml

Fresh APS “crackles” when water is added. If yours does not, replace it with
fresh stock. Prepare just prior to use.

Sequencing gel acrylamide solution

(6%, 100ml)

Acrylamide

stock

solution,

40%

6%

15ml

Urea

(FW

60)

7M

42g

TBE,

5X 1X

20ml

Add

deionized

H

2

O

to

100.0ml

Dissolve

Filter,

dearate,

and

add:

APS,

10%

solution

1ml

TEMED

60ul

Pour

gel

immediately.

Advertising
This manual is related to the following products: