2 dc-powered sonartractm, Figure 41 transmitter power connection -11, Warning – CiDRA SONARtrac GVF-100 User Manual

Page 50: Caution

Advertising
background image

Copyright © 2006 CiDRA Corporation

Page 8-11

20675-01 Rev 02

Feed electrical power wires through the fitting. Referring to the
following figure, attach the ground wire (green) to the Ground ( )
terminal , hot (black - U.S., brown - Eur) to the L (+) terminal, and
neutral (white – U.S., blue – Eur) to the N (-) terminal.

Figure 41

Transmitter Power Connection

8.4.3.2 DC-Powered

SONARtrac

TM

Any voltage within the range of 18 – 36 VDC can be applied to the DC
version of the SONARtrac

TM

. Power cables of size 18-gauge

(minimum) to 10-gauge (maximum), with a ground conductor, are
required.

WARNING

For Class I, Division 2 installations, a non-current carrying safety
ground attached to the ground terminal on the input power
terminal block is required, and Power Entry and Inputs/Outputs
must be installed in accordance with Article 501.10(B)(1) of the
National Electrical Code ANSI/NFPA 70:2005.

CAUTION

Always use a non-current carrying safety ground attached to the
ground terminal on the input power terminal block. Failure to do
so could result in poor system operation.

Feed electrical power wires through the fitting. Referring to the
previous figure, attach the ground wire to the Ground ( ) terminal,
DC+ to the L (+) terminal, and DC- to the N (-) terminal.

Line In (+)

Neutral (-)

Ground (

)

Advertising
This manual is related to the following products: