DFX-7 Ultrasonic Flaw Detector

59

2) Press the LEFT and RIGHT arrow keys to position the start of GATE1 to

trigger on the waveform.

3) Optionally, press the MEAS key multiple times to move right and the ESC

key multiple times to move left, until the WIDTH1 cell is highlighted.

4) Press the LEFT and RIGHT arrow keys to adjust the size or width of

GATE1.

Note: When the signal is in the gate, the height indicator will detect and display

the % full screen height.

Setting the Level

1) Press the MEAS key multiple times to move right and the ESC key multiple

times to move left, until the AWS BOX: cell is highlighted.

2) Press the LEFT and RIGHT arrow keys to position the cursor box over the

waveform.

3) Press the ENTER key to set the REFGN (reference gain).

Note: This process can be repeated and reset multiple times, or until the SET

REF GAIN feature is toggled OFF.

4) Press the MENU key once to activate the menu items tab. Press the MENU

key multiple times to tab right, and the ESC key multiple times to tab left,

until the AWS menu is highlighted and displaying the submenu items.

5) Press the UP and DOWN arrow keys until SET REF GAIN is highlighted,

followed by pressing the LEFT and RIGHT arrow keys to toggle OFF.

6) Finally, press the MEAS key to return to the main measurement screen,

followed by pressing the OK key to display the AWS screen.