DFX-7 Ultrasonic Flaw Detector

63

2) Use the UP and DOWN arrow keys to scroll through the sub menu items

until DAC is highlighted.

3) Press the LEFT and RIGHT arrow keys to toggle the DAC feature ON or

AUDIBLE. The menu items should now become active and adjustable, as

illustrated above.

Activating Draw Mode

1) Press the UP and DOWN arrow keys until DRAW is highlighted, followed by

pressing the LEFT and RIGHT arrow keys until ON is displayed.

Quantity Displayed

1) Press the UP and DOWN arrow keys until MEAS is highlighted, followed by

pressing the LEFT and RIGHT arrow keys to toggle and select dB, %FSH,

or %DAC.

Selecting the Curve(s)

1) Press the UP and DOWN arrow keys until CURVE is highlighted, followed

by pressing the LEFT and RIGHT arrow keys to toggle and select DAC, -2, -

4, -6, -10, (-6/-12), or (-6/-14) curve options. The CURVE option can be

changed at any time after the DAC curve has been created.

Setting the Curve(s) Trigger

1) Press the UP and DOWN arrow keys until TRIGGER is highlighted, followed

by pressing the LEFT and RIGHT arrow keys to toggle and select which

curve or gate to use as the alarm trigger. The options displayed will vary

according to which CURVE option has been selected above: DAC, -2, -4, -6,

-10, (-6/-12), or (-6/-14) curve options. The Trigger option can be changed

at any time after the DAC curve has been created.

Selecting the Fit