3) Alternatively, press the ENTER key to display the Digits Edit Box.

4) Press the UP and DOWN arrow keys to scroll the highlighted value.

5) Press the LEFT and RIGHT arrow keys to scroll the digit locations.

6) Repeat steps 4 & 5 until the GATE value is correctly displayed.

7) Press the OK key to return to the measurement screen, or ESC to

cancel entering the GATE value.

The user can also access and adjust the gate from the tabbed menus. However,
this method is more tedious than making the adjustments using the Hot Menus.
The procedure using the tabbed menus is outlined below:

Adjusting the Gate using the Tabbed menus

1) Press the MENU key once to activate the menu items tab. Press the

MENU key multiple times to tab right and the ESC key multiple times to

tab left until the AUTO menu is highlighted and displaying the submenu

items.