Performance Specifications
Minimum and Maximum Operating Conditions
Condition
Minimum
Maximum
Inlet Pressure
40 psi (2.76 bar)
80 psi (5.52 bar)
Inlet Temperature
40°F (4.44°C)
100°F (37.78°C)
Inlet TDS
50 PPM
1,000 PPM
Inlet Hardness
0 mg/L (0 grain)
171 mg/L (10 grain)
Inlet Chlorine
0 mg/L
1.0 mg/L
Inlet Iron
0 mg/L
0.1 mg/L
Inlet Manganese
0 mg/L
0.05 mg/L
Inlet pH
4
10
Inlet Turbidity
0
1 NTU
12
Silica 0 mg/L 3 mg/L