5 installation and removal, Installation of the bus module 5.1.1, Establish connections – INFICON BM1000 Bus module User Manual

Page 10: 5installation and removal, 1 installation of the bus module, 1 establish connections

Advertising
background image

10

Installation and removal

5

Installation and removal

5.1

Installation of the bus module

DIN-TS35 top hat rail

1

Hook unit on top hat rail at bottom.

2

Press unit onto top hat rail at top.

5.1.1

Establish connections

Connection of bus module with MSB box

The bus module communicates via a data cable with the mass spectrometer module
and is supplied with voltage by means of the data cable.

Data cables from INFICON

1

Connect bus module (connection L

D

) via data cable with MSB box

(connection I/O).

Information If you disconnect the data cable from the connection L

D

of the bus

module during operation, the communication between the bus module
and the MS module will be interrupted.

2

Connect the bus module with the external controller by means of the built-in
Fieldbus module.

Communication with Fieldbus master

► For information on establishing communication with the Fieldbus master, see

Interface protocols, jira54.

Fig: 2

Mount bus module on DIN

-

TS35 top hat rail