1.4 Spare part list
Part no.
Battery
591-294
Earphones 591-443
Carrying case
591-304
Charger AC
(591-300) 591-301 SV, 591-302 USA, 591-303 UK
Charger car
591-361
Hand Probe H21
590-200
Cable C21
590-150
Circuit board
591-293
1.5 Function check
Action
Result
Connect charger The instrument starts
Disconnect The instruments turns off
Change sensitivety The LED’s indicates corresponding sensitivity
Adjust the volume The volume should change
Press “Battery Check” button The LED should indicate Battery status
Disconnect Hand Probe The ON LED should flash red and an audio
signal appear
1.6 Sensitivity check
Action
Result
Start XRS9012 The instruments starts and indicates sensitivity 5
Adjust the sensistivity to 10 The no. 10 LED is flashing
Apply 10ppm Hydrogen Signal should indicate: minimum LED 1