2 push-buttons, 3 leds, 4 ports and connections – INFICON Sensistor ISH2000 Hydrogen Leak Detector User Manual

Page 11: Push-buttons, Leds, Ports and connections

Advertising
background image

Hydrogen leak detector Sensistor ISH2000

-

Operating Manual

OM-Sensistor ISH2000-EN-201304

11

EN

3.2 Push-buttons

The functions of the push-buttons are shown at the lower edge of the display. In this
manual the buttons are numbered, from left to right, 1, 2, 3, and 4. The push-buttons
are used to:
•

Change from one menu item to another using the

<

and

>

buttons.

•

Press

Enter

to move down to the nearest sub-menu.

•

Press

Save

to save the set value.

•

Press

Undo

to restore the previously set value.

•

Press

Esc

to move up to the nearest higher level(s).

•

Shortcut to reach functions as Volume Sensitivity (Sens), Calibration (Cal) and
Menu.

3.3 LEDs

The two LEDs indicate the status of the instrument as follows:
•

Green flashing slowly, during warming up phase.

•

Green fixed light indicates that instrument is ready and hydrogen signal is below
Reject Level limit.

•

Periodically flashing green or red calls for calibration.

•

Alternating red and green light while the instrument establishes probe
communication.

•

Red fixed light together with

Reject

on display means the instrument has detected a

leak larger than the set Reject Level limit.

•

Red flashing rapidly, check message on screen. (See “Trouble-shooting” on
page 49.)

3.4 Ports and connections

The ports and connections are shown in Figure 3-2 on the next page.

NOTICE!

Always connect all four wires to the Power connector to
24VDC in order to operate.