Warranty and Liability-Limitation

INFICON warrants your D-TEK Select Refrigerant Leak Detector to be free from
defects of materials or workmanship for two years from the date of purchase.
INFICON does not warrant items that deteriorate under normal use, including
power stick, infrared absorption element and filters. In addition, INFICON does not
warrant any instrument that has been subjected to misuse, negligence, or
accident, or has been repaired or altered by anyone other than INFICON.
INFICON liability is limited to instruments returned to INFICON, transportation
prepaid, not later than thirty (30) days after the warranty period expires, and which
INFICON judges to have malfunctioned because of defective materials or
workmanship. INFICON liability is limited to, at its option, repairing or replacing the
defective instrument or part.
This warranty is in lieu of all other warranties, express or implied, whether of
MERCHANTABILITY or of FITNESS FOR A PARTICULAR PURPOSE or
otherwise. All such other warranties are expressly disclaimed. INFICON shall
have no liability in excess of the price paid to INFICON for the instrument plus
return transportation charges prepaid. INFICON shall have no liability for any
incidental or consequential damages. All such liabilities are EXCLUDED.

3) After warm up
sequence the green,
the last yellow and
the amber charging
LEDs flash.

3a) Battery needs
recharging.

3a) Charge battery for 10-12
hours.

3b) Power stick has
failed.

3b) Replace power stick with
part #712-700-G1.

4) Pump is not
working.

4a) Pump has failed.

4a) See 2b.

5) Unit does not
power up. Hi/Lo
indicators flash in a
rapid, alternating
manner when AC
adapter is
connected.

5a) Power
stick/battery has
been connected in
reverse.

5a) Reverse the battery's
connection and power unit
up again.

Problem

Cause

Remedy