INFICON BPG402-SE ATM to Ultra-High Vacuum Gauge (EtherCAT) User Manual
Page 15
tira93e1 (2014-04) BPG402-SE.cp
15
Index
SI
DataType
NV Access PM Name
8015
0x01 BOOL
x
RW
Filament User Mode
Module 2
0x02 BOOL
x
RW
Emission User Mode
Module 2
0x11 BYTE
x
RW
Active Filament Module 2
0x13 REAL
x
RW
Emission Current Module 2
Defines whether the filament selection is static as defined by "Active Filament" =
manual mode or the active filament is changed always when the filament is
switched on = automatic mode
Filament User Mode Module 4
0
1
Automatic
Manual
Defines whether the emission is switched on automatically by a different sensor
(automatic) or manually by the user (manual)
Filament User Mode Module 4
0
1
Automatic
Manual
Active Sensor filament
Active Filament Module 4
Bit 0
Bit 1
Bit 1…7
Filament 1
Filament 2
0
Emission current in milliamps
4.5 Configuration Area
(0x8000…0x8FFF)
4.5.1 Configuration Hot
Cathode Ion
Subindex 0x01
Subindex 0x02
Subindex 0x11
Subindex 0x13