General Information

The BPG400-SP gauge has a fieldbus interface
that conforms to the Profibus DPV1 standard
(

→ [9]).

Via this interface, the following and further data
are exchanged in the standardized Profibus
protocol (

→ [1], [2]):

• Pressure reading
• Pressure unit (Torr, mbar, Pa)
• Degas function
• Gauge adjustment
• Status and error messages

Two adjustable switching functions are integrated in the
gauge. The corresponding relay contacts are available at the
sensor cable connector.

The basic sensor and sensor electronics of the BPG400-SP
type are the same as in the standard BPG400 (

→ [1], [2]).

Technical Data BPG400-SP

General technical data of the sensor and sensor
electronics

→ [1], [2].

Fieldbus Interface

Fieldbus name

Profibus

Standard applied

→ [9]

Communication protocol,
data format

→ [4], [9]

Interface, physical

RS485

Profibus Parameters

Data rate

≤12 MBaud
(

→ [4], [9])

Node address

00 … 7D

hex

(0 … 125

dec

)

Profibus connection

D-Sub, 9 pins, female

Cable

Shielded special Profibus
cable (

→ [9], [7])

Cable length, system wiring

According to Profibus
specifications (

→ [9],

[7])

Supply Voltages

The power consumption of the BPG400-SP is higher
than that of the standard BPG400.

Supply voltage at sensor cable
connector, Pin 8

+24 VDC (+20 … 28 V)

Power consumption

<18 W

Sensor Cable Connection

→ "Technical Data, Sensor Cable Connection" of the gauge
BPG400-SD (identical).

Dimensions

Housing and vacuum connection

→ [1], [2]

Weight

353-505
353-506

≈425 g

≈685 g

Electrical Connection

Make sure the vacuum connection is properly made
(

→ [1], [2], "Vacuum Connection").

Sensor Cable Connection

n

If no sensor cable is available, make one according to
the diagram shown in the BPG400-SD section
(identical).

o

Connect the sensor cable to the gauge and secure the
sensor cable connector using the lock screws.

Profibus Cable Connection

n

If no Profibus cable is available, make one according to
the following indications:

1

5

6

9

D-Sub, 9 pins, male
soldering side

Pin 1

not connected

Pin 2

not connected

Pin 3

RxD/TxD-P

Pin 4

CNTR-P

1)

Pin 5

DGND

2)

Pin 6

VP

2)

Pin 7

not connected

Pin 8

RxD/TxD-N

Pin 9

not connected

1)

Only to be connected if an optical link module is
used.

2)

Only required as line termination for devices at both
ends of bus cable (

→ [9]).

o

Connect the Profibus cable to the gauge and secure
the Profibus cable connector using the lock screws.

The gauge can now be put into operation.

Operation

Caution

Caution: data transmission errors
If the gauge is operated via an RS232 interface
and the DeviceNet at the same time, data trans-
mission errors may occur.
The gauge must not be operated via an RS232
interface and the DeviceNet at the same time.

Operating Software

For operating the gauge via the Profibus network, prior in-
stallation of the gauge specific GSD file is required. This
software can be downloaded via internet (

→ [5]).

Note Address Setting

The node address (0 … 125

dec

) is set in

hexadecimal form (00 … 7D

hex

) via the

"ADDRESS", "MSD", and "LSD" switches.
The node address is polled by the firmware
when the gauge is switched on. If the
setting deviates from the stored value, the
new value is taken over into the NVRAM. If
a value >125

dec

(>7D

hex

) is entered, the

node address setting currently stored in the
device remains valid but it can now be
defined via Profibus ("Set slave Address",

→ [4]).

Adjusting the Gauge

→ "Adjustment and settings".

Adjusting the Switching Functions

→ "Adjustment and settings".

BPG400-SR (RS485)

General Information

The BPG400-SR gauge has a RS485 fieldbus
interface.
Via this interface, the following and further data
are exchanged (

→ [1], [2]):

• Pressure reading
• Pressure unit (Torr, mbar, Pa)
• Degas function
• Gauge adjustment
• Status and error messages
• Threshold values of switching functions

Two adjustable switching functions are integrated in the
gauge. The corresponding relay contacts are available at the
sensor cable connector. Additionally, the Relay contact of
switchingfunction A is also available at the interface
connector.

The basic sensor and sensor electronics of the BPG400-SR
type are the same as in the standard BPG400 (

→ [1], [2]).

Technical Data BPG400-SR

General technical data of the sensor and sensor electronics

→ [1], [2].

Fieldbus Interface

Name

RS485

Communication protocol,
data format

→ [2]

RS485 parameters

Data rate

300 … 28'800 Baud

Device address

00 … 7F

hex

, (0 … 127

dec

)

RS485 connection

D-Sub, 9 pins, male

Cable

1 twisted pair, shielded
(

→ [2])

Supply Voltages

The power consumption of the BPG400-SR is higher
than that of the standard BPG400.

Supply voltage at sensor cable
connector, Pin 8

+24 VDC (+20 … 28 V)

Power consumption

<18 W

Sensor Cable Connection

→ "Technical Data, Sensor Cable Connection" of the gauge
BPG400-SD (identical).

Dimensions

Housing and vacuum connection

→ [1], [2]

Weight

353-509
353-513

≈425 g

≈685 g

Electrical Connection

Make sure the vacuum connection is properly made
(

→ [1], [2], "Vacuum Connection").

Sensor Cable Connection

n

If no sensor cable is available, make one according to
the diagram shown in the BPG400-SD section
(identical).

o

Connect the sensor cable to the gauge and secure the
sensor cable connector using the lock screws.

RS485 Cable Connection

n

If no RS485 cable is available, make one according to
the following indications: