tina80e1 (2012-06)

21

Evacuate the gauge to a pressure according to the table

below:

F.S.

Recommended final pressure for

zero adjustment

1100 mbar
1000 Torr/mbar
200 Torr/mbar
100 Torr/mbar
50 Torr/mbar
20 Torr/mbar
10 Torr/mbar


<5×10

-2

Torr

<10

-2

Torr

<5×10

-3

Torr

<2.5×10

-3

Torr

<10

-3

Torr

<5×10

-4

Torr

<6.65×10

0

Pa

<6.65×10

0

Pa

<1.33×10

-0

Pa

<6.65×10

-1

Pa

<3.33×10

-1

Pa

<1.33×10

-1

Pa

<6.65×10

-2

Pa

<5×10

-2

mbar

<5×10

-2

mbar

<10

-2

mbar

<5×10

-3

mbar

<2.5×10

-3

mbar

<10

-3

mbar

<5×10

-4

mbar

If the final pressure in the gauge is too high for zero ad-
justment (>25% of the F.S.), the zero cannot be reached.

Briefly press the button with a pin (max. ø1.1 mm). The

zero adjustment runs automatically (duration ≤8 s).

Press the button briefly

max. ø1.1 mm

After zero adjustment the gauge automatically returns to

measurement mode.