JULABO F38-ME Beer Forcing Test Refrigerated/Heating Circulating Bath User Manual
Page 34

Menu functions
34
9.2.2. STANDARD
Program
Menu level 1
1 program
10 sections
The integrated programmer permits fast and easy programming of setpoint
temperature sequences. This temperature sequence is called program. A
program is composed of individual sections (STEP). The sections are defined
by duration (TSLICE) and target temperature. The target temperature is the
setpoint (SETPNT), which is achieved at the end of a section. The programmer
calculates the temperature ramp from the difference in time and temperature.
STEP
(Nr.)
1 2 4 6 8
SETPNT
(°C) 100 180 180 75 75
TIME (hh.mm)
00:20 00.10 00:20 00:50 00:20
Graph1
25
50
10 20 30 40 50 60 10 20 30 40 50 60 10 [min]
75
100
150
200
°C
t
Step 1
2
4
6
8
Sections without set value and time are skipped. They can be defined
retroactively and the integrated into the program.