JULABO F38-ME Beer Forcing Test Refrigerated/Heating Circulating Bath User Manual
Page 38

Menu functions
38
Program sequence:
Example forcing test 60 °C
As factory-setting the „Forcing Test 60 °C“ is stored
Measuring cycle: RUNS
1 repetition
Warm phase
W-TEMP
60 °C
Duration
W-TIME
24 h
Cool phase
C-TEMP
0 °C
Duration C-TIME
24
h
In the course of the program the VFD-DISPLAY shows messages
concerning the process of the forcing test in line 2 and 3.
In line 3 two messages are shown alternately.
Message 1 always indicates the program status (phase of the forcing test),
message 2 indicates the status >RUN< (Test runs).
Message 2 can be changed by the editing keys
.The remaining
time of a phase or >E ------< can still be chosen.
(Message >E ------<, temperature value of the external temperature sensor
is not important for the forcing test.)
Action
diagram:
1.
Set measuring temperature :
Select in the MENU
e.g. SETPNT 1 and set 20 °C.
(see page 22)
2. Start forcing test (see page 29)