Leica knife holder CE /CN

1.

Operation

26

27

Fig. 14

Clearance angle adjustment - knife holder CN

Prior to setting the clearance angle, the knife
edge must be located exactly in the spin axis of
the knife holder.

First, set a clearance angle of ‘0’. To do so, loos-
en lever (23, see Fig. 11) and bring the index
marker (26) in line with the figure ‘0’ on the clear-
ance angle scale (27). Retighten the lever (23). If
the sectioning results obtained are not satisfac-
tory, slowly increase the clearance angle in 1°
steps, until the desired section quality is
achieved.

28

29

Fig. 15

Parallel adjustment - knife holder CE
(Adjusting the anti-roll system)

This type of adjustment has to be done each time
after exchanging either the entire anti-roll sys-
tem or the change frame.
Loosen Allen screw (28) and then, via the axle
extension (29), adjust the upper edge of the glass
anti-roll plate parallel to the knife edge. Retighten
Allen screw (28).

Parallel adjustment - knife holder CN
(Adjusting the anti-roll system)

After loosening the two screws (30) the anti-roll
system can be adjusted parallel to the knife edge.
After carrying out the adjustment, retighten the
screws (30).

Fig. 16

30

31

32