Leica Biosystems Bond Oracle HER2 IHC System Instructions for Use TA9145 EN-CE-Rev_E 18/06/2013

English

and/or CLSI (formerly NCCLS) Quality Assurance for Immunocytochemistry, Approved

Guideline (12). These quality control procedures should be repeated for each new antibody lot,

or whenever there is a change in assay parameters. Human invasive (infiltrating) ductal breast

carcinoma with known HER2 oncoprotein staining intensities from 0 to 3+ and other suitably

negative tissues are appropriate for assay verification.

Interpretation of Staining

For the determination of HER2 oncoprotein expression, only membrane staining pattern and

intensity should be evaluated using the scale presented in Table 4. A pathologist using a bright-

field microscope should perform slide evaluation. For evaluation of the immunohistochemical

staining and scoring, an objective of 10x magnification is appropriate. The use of 20–40x

objective magnification should be used in the confirmation of the score. Cytoplasmic staining

should be considered as nonspecific staining and is not to be included in the assessment of

membrane staining intensity (14). To aid in the differentiation of 0, 1+, 2+, and 3+ staining, refer

to the Bond Oracle HER2 IHC System Interpretation Guide for representative images of the

staining intensities. Only specimens from patients with invasive breast carcinoma should be

scored. In cases with carcinoma in situ and invasive carcinoma in the same specimen, only the

invasive component should be scored.

Immunohistochemical Staining Pattern

Score

Assessment

No staining is observed or membrane staining is observed in less
than 10% of the tumor cells.

0

Negative

Faint/barely perceptible membrane staining is detected in more
than 10% of the tumor cells. The cells are only stained in part of
their membrane.

1+

Negative

Weak to moderate complete membrane staining is observed in
more than 10% of the tumor cells.

2+

Equivocal
(Weakly Positive)

Strong complete membrane staining is observed in more than
10% of the tumor cells.

3+

Strongly Positive

Table 4. Interpretation of HER2 staining

Bond Oracle HER2 IHC System staining results are interpreted as negative for HER2

oncoprotein expression with scores of 0 and 1+ staining intensity, equivocal (weakly positive)

with a score of 2+ staining intensity, and strongly positive with a score of 3+ staining intensity.

Bond Oracle HER2 IHC System is not intended to provide prognostic information to the patient

and/or physician and has not been validated for that purpose. For each staining assessment,

slides should be examined in the order presented below to determine the validity of the staining

run and enable semi-quantitative assessment of the staining intensity of the sample tissue.