Calculating the Total Magnification and Field of View Diameter

Parameter

MO

Magnification of objective

ME

Magnification of eyepiece

z

Magnification changer position

q

Tube factor, e.g. 1.5× for coaxial incident light, 1.6× for 45°
ErgoTube™

r

Factor 1.25× if the planachromatic and planapochromatic
objectives of the MZ125/MZ16 are used on the MS5, MZ6,
MZ75 or MZ95

NFOV Field number of the eyepiece. Field numbers are printed on the

eyepieces: 10Ч = 21, 16Ч = 14, 25Ч = 9.5, 40Ч = 6.

Example

MO

1× objective

ME

25×/9.5 eyepiece

z

Zoom position 4

q

Coaxial incident light 1.5×, tube factor

r

Factor 1.25×

Calculation example: magnification in the binocular tube

MTOT VIS = MO × ME × z × q × r

or

1 Ч 25 Ч 4 Ч 1.5 Ч 1.25 = 187.5Ч

Calculation example: field of view diameter in the specimen

∅ OF:

NFOV

MO × z ×q Чr