4 basic equipment – Memmert HPP 108/749 Constant climate chamber (Generation 2003) User Manual

Page 11

Advertising
background image

11

Design and description

Range A:

In this range, temperature and humidity can be combined as you please, without resulting in

any significant condensation. If there are extreme ambient conditions, the working range may

be restricted.

Range B:

If the specified range is exceeded upwards, e.g. 80 % rh at 60°C, the hot steam fed in will im-

mediately condense, due to the dew-point, at the coldest point in the appliance.

Range C:

At low temperatures and low relative air humidity, the effective range is heavily dependent on

the degree of humidity of the chamber load.

2.4

Basic equipment

►

Low-noise, long-life and energy-saving Peltier cooling and heating technology (in heat-

ing mode, part of the energy required is extracted from the surroundings = heat pump

principle)

►

Electronic fuzzy-supported PID process controller with permanent performance adjust-

ments and time-saving self-diagnosis system to quickly locate errors (see pages 38 and 45)

►

Alphanumeric text display

►

Control of the appliance and logging the actual values with a preformatted empty

MEMoryCard XL (storage capacity 32 kByte, reprogrammable for up to 40 ramp segments

and in addition 135 hours of log memory for temperature and humidity with a scanning

interval of 1 minute)

►

Programme sequence control for up to 40 ramp segments

►

Integrated week time switch with group function (e.g. all working days)

►

Retracting push-turn control for simple operation

►

Optical alarm display

►

Integrated horn as alarm if limit values are crossed, as acoustic signal when programme

ends and to confirm input (button click)

►

Digital monitoring control for overtemperature, undertemperature and automatic tem-

perature monitor (ASF)

►

Two separate Pt100 temperature sensors DIN class A in a 4-wire circuit for control and

monitoring

►

Parallel printer port (PCL3-compatible)

►

USB interface for computer-based temperature control programmes and to read out the

controller’s internal log memory

►

MEMMERT software "Celsius" for remote control of the chamber via computer (see page

37) and to read out the internal controller log memory

►

As an optional accessory, external card reader for MEMory-Card XL for connection to the

PC-USB interface, printer cable (parallel, shielded)- 25-pin

►

Language setting (German, English, French, Spanish, Italian)

►

Capacitive humidity sensor

►

Active humidity control guarantees that setpoint humidity is quickly achieved, with short

recovery times, and avoiding condensation formation