3 electrical connections, 4 switching on – Memmert HPP 108/749 Constant climate chamber (Generation 2003) User Manual

Page 21

Advertising
background image

21

Putting into Operation

4.3 Electrical

connections

Caution:

Observe the country-specific regulations when making connections (e.g. in Germany

DIN VDE 0100 with residual current device (RCD)). Remember the connected loads and

power values (see nameplate and also chapter " Technical data " on page 14 ).

The constant climate chamber is intended for operation on an electrical power system with

a system impedance Z

max

at the point of transfer (service line) of a maximum of 0.292 ohm.

The operator must ensure that the constant climate chamber is operated only on an electrical

power system that meets these requirements. If necessary, you can ask your local energy sup-

ply company what the system impedance is.
Only appliances may be connected to external connections whose interfaces comply with the

requirements for safety extra-low voltage (e.g. computer, printer).

4.4

Switching on

To switch on the appliance, press the main

switch on the front of the appliance ( Fig. 11 ).

Caution:

The first time the appliance is oper-

ated, it must not be left unattended

until it has reached the steady state.

Fig. 11

Switching on the constant climate

chamber

Advertising