8

Normal operation

PRINT

SETUP

In this operating mode the oven operates continuously. The settings for operating the oven can be selected.
The settings act directly on the operation of the oven.

MIN

MAX

AUTO

°C

By rotating the push/turn control the following parameters can be selected and can be altered as described
in the

Section „Setting the parameters“

:

Temperature setpoint
Range: 20 °C up to nominal temperature
(details see label)

˚C

Fan speed
Range: 0 to 100%
in 10% steps.

Air valve
Range: 0 to 100%
in 10% steps.

Temperature monitor
Adjustment range: MIN MAX AUTO

(see Section “Temperature monitor“)

MIN

MAX

AUTO

°C