Programming example steriliser, Select operating mode “programme, Edit programme – Memmert IFP 400 - 800 Incubators (Generation 2003) User Manual

Page 44: Immediate programme start, Select duration of first ramp segment, Set temperature of first ramp segment, Set fan speed for first ramp segment

Advertising
background image

PERFECT

page 44

Programming example steriliser

The steriliser (SFP600) has to sterilise at a temperature of 180°C and a medium quantity of load for one
hour and 30 minutes. By setting a cooling time the load can only be removed after it has cooled down.

sterilisation time

heating time

180°C

cooling time

°C

t=time

1. Select operating mode “Programme“

After holding down the SET key (approx. 3 sec) the current operating
mode is flashing. Hold down the SET key and select operating mode
“Programme“ using the push/turn control.
After the SET key has been released, the controller is in the “Programme”
operating mode.

PRINT

SETUP

2. Edit programme

Hold down the SET key and turn the push/turn control to select “EDIT”.
After the SET key has been released, the controller is in the programme
writing mode.

3. Immediate programme start

Hold down the SET key and turn the push/turn control to select INSTANT

START.

4. Select duration of first ramp segment

Turn the push/turn control to select the time indication.
Hold down the SET key and set the time 00:01 using the push/turn
control.

h

5. Set temperature of first ramp segment

Turn the push/turn control clockwise until the temperature display is
flashing.
Hold down the SET key and set the required temperature setpoint of

180 °C using the push/turn control.

˚C

6. Set fan speed for first ramp segment

Turn the push/turn control clockwise until the fan symbol is flashing.
Hold down the SET key and set the fan speed to 100% using the
push/turn control.

The fan motor is running at
maximum speed.

Advertising