Charger connection, Series 248 actuator product information, How to precharge the static support operation – MTS Series 248 Actuators User Manual
Page 40: Caution

Series 248 Actuator Product Information
40
How to Precharge the Static Support
Operation
Charger Connection
Rapid flow rates with pressure differentials of more than 2.1 MPa (300 psi)
across the input check valve can occur when you open the regulator shut-
off valve.
Damage to the valve seals can result from rapid flow rates.
Open the regulator shut-off valve only far enough to permit a gradual transfer of
gas.
Extension Hose
r
Bleed Valve
Input
Check Valve
Open
Close
Nitrogen Supply Hose
Regulator Output
Pressure Gage
THESE ITEMS SUPPLIED BY USER
Nitrogen Bottle
Pressure Gage
Regulator
Shut-Off Valve
Regulator Output
Pressure Valve
Nitrogen
Bottle Valve
Chuck Valve
Gage Protector
(factory set to limit
pressure to the gage
to approximately
1.4 MPa (200 psi)
High Pressure Gage
0-21 MP (0-3000 psi )
Poppet Type Valve
To Open and Close
Use Locknut
Open
Valve
Stem
Nitrogen
Bottle
Low Pressure Gage
0-2.1 MPa (0-300 psi)
Locknut
Close
Open
Close
Open
Close
Open
Close
CAUTION