J6 and j7 monitor, J8 and j9 dc conditioner – MTS ReNew Technical Reference User Manual
Page 31

ReNew Technical Reference
Installation
31
J6 and J7
Monitor
Two Monitor connectors are provided. There are several possible uses for analog
monitor outputs: external data acquisition, tuning, troubleshooting, etc. For
tuning, it is desirable to monitor command and feedback, or command and error,
simultaneously while changing the controller parameters. Therefore, two monitor
outputs are provided. Specifics for these connectors are:
•
Analog +/-10.5 V
•
Calibrated to +/-10 V
•
16-bit resolution minimum
•
BNC connectors
Note
The load cell is mounted to the crosshead and its wiring is internal to the
frame, thus its connector is not on the rear panel.
J8 and J9 DC
Conditioner
Two DC Conditioner connectors are provided. The two application specific
transducers might be biaxial strain gage base extensometers. With external
completion resistors, the DC conditioners could be used with quarter bridge
strain gages. Pin assignment is as follows:
Pin
Signal
1
TEDS data
2
EX+
3
EX-
4
FB-
5
RCAL1 (FBR+)
6
RCAL2 (FBR-)
7
FB+
8
EXS-
9
EXS+
10
TEDS ground