ReNew Technical Reference

42

Troubleshooting Guide

Crosshead will not stop

•

Servo is improperly connected.

•

Improper machine parameters. Contact an MTS Applications Engineer for
further assistance.

•

Controller error. Contact an MTS Applications Engineer for further
assistance.

Limit is always active —

cannot clear

•

Proximity limit switch on the crosshead is obstructed.

•

Limit switch is defective. Contact an MTS Applications Engineer for further
assistance.

•

Frame controller limit switch circuitry is malfunctioning. Contact an MTS
Applications Engineer for further assistance.

Limit switches not

functioning

•

Defective limit switch. Same as above. DANGER — DO NOT RUN
MACHINE!
Contact an MTS Applications Engineer for further assistance.

Will not pull rated

maximum load

•

Defective servocontroller. Contact an MTS Applications Engineer for
further assistance.

•

Load cell is not calibrated.

Improper machine

parameters

•

Contact an MTS Applications Engineer for further assistance.

Load cell will not

calibrate

•

Load cell is not connected.

•

Defective load cell.

•

You are attempting to automatically calibrate a non-MTS load cell. These
must be manually calibrated.

•

Improper calibration information being used. Check the load cell
information in the testing software.

Load cell amplifier

defective

•

Contact an MTS Applications Engineer for further assistance.

Load plateaus

•

Defective load cell.

Unusual noise

•

Grinding sound indicating that a moving part is rubbing. Lubrication is
needed. Inspect belt and screw bearings. Contact an MTS Applications
Engineer for further assistance.

•

Debris caught in the cooling fan.