Calibration

23

The following table shows the proper conditioner output for each clip-
on gage displacement of the preceding example.

.

How to Use a Calibration Block

MTS recommends using a calibration block to establish the initial gain
setting of the conditioner. The calibration block is a metal block with
knife edges mounted to it or machined in it. Using multiple knife edges
allows you to adjust the gain for a variety of specific displacements.

Use this procedure as a guideline to calibrate other clip-on gages.

1. Set up the controller.

Your test controller must be configured to use the clip-on gage
signal.

A. Connect the clip-on gage to your controller.

B. Monitor the excitation voltage.

C. Monitor the clip-on gage signal.

D. Prepare to use the calibration controls on your controller.

E.

Adjust the excitation control to set the excitation to +6 V DC
or the voltage specified in the clip-on gage documentation.

2. Zero the conditioner output.

% F

ULL

S

CALE

T

RAVEL

O

UTPUT

100%

3 mm

+10 V DC

80%

2.4 mm

+8 V DC

60%

1.8 mm

+6 V DC

40%

1.2 mm

+4 V DC

20%

0.6 mm

+2 V DC

Zero

0 mm

0 V DC

-20%

-0.6 mm

-2 V DC