How to connect the cable, How to connect the cable 30 – MTS Clip-On Gages User Manual

Page 30

Advertising
background image

Installation

30

Knife edges must be added to the specimen.

Depending on the model of the clip-on gage and the specimen
standard, knife edges are either machined into the specimen or
attached to the specimen.

If the knife edges are attached to the specimen, use the
alignment fixture to establish the edge length and knife edge
alignment.

How to Connect the Cable

The clip-on gage cable connects to a controller via an extension cable.
The clip-on gage assembly includes a small cable and connector. An
extension cable is installed between the clip-on gage connector and the
system controller. Ensure that the clip-on gage is connected to the
appropriate controller connector. The controller connector must be
associated with a DC conditioning circuit. The following figure shows
the circuitry and connector pin assignments from the clip-on gage.

GL XXX

60°

The knife edges are mounted
onto the specimen

Knife
Alignment
Fixture

The knife edges
are machined into
the specimen.

Knife
Edges

The shape of the knife edges
can vary between models.

Grip Mounting Holes

Crack Starter Notch

Advertising