13 • Operation

Model GP-01 Operator’s Manual

Alarms

This section covers alarm indications. It also tells you how to reset the
Model GP-01 after an alarm has occurred and how to respond to an alarm
condition.

Alarm Indications

The Model GP-01 will sound an alarm, flash its alarm lights, and vibrate
when the target gas concentration rises above the low alarm point. The
Model GP-01 also sounds an alarm, flashes its alarm lights, and vibrates
when the high alarm point is reached. In addition, the Model GP-01 has a
low battery warning, a dead battery alarm, an over range alarm, a sensor
failure alarm, and a system failure alarm. See Table 4 below for a
description of each alarm indication.

Table 4: Alarm Types and Indications

Alarm Type

LCD Indications

Other Indications

Low Alarm

Concentration of gas
rises above the low
alarm point.

• Gas reading

flashes.

• Back light turns on.

• Pulsing tone occurs

once per second.

• Unit vibrates once

per second.

• Alarm lights flash

once per second.

High Alarm

Concentration of gas
rises above the high
alarm point.

• Gas reading

flashes.

• Back light turns on.

• Pulsing tone occurs

twice per second.

• Unit vibrates twice

per second.

• Alarm light flashes

twice per second.

Over Range
Concentration of gas
rises above the
measuring limit of the
Model GP-01. (Or
there could be a
problem with the unit.)

• Gas reading

replaced by blinking
brackets.

• Back light turns on.

• Pulsing tone occurs

once per second.

• Unit vibrates once

per second.

• Alarm light flashes.

Low Battery Warning

• Last remaining bar

on the right in
battery icon
flashes.

• None