Model GP-01 Operator’s Manual

Operation • 14

Resetting Gas Alarms

To reset a gas alarm, press the POWER/MODE button after the gas
reading falls below the low alarm point.

NOTE: If the gas concentration has returned to normal or has fallen below

the low alarm point, the alarm indications will continue until you
reset the alarm using the POWER/MODE button.

Responding to Alarms

This section describes response to gas, over range, battery, sensor failure,
and system failure alarms.

Responding to Gas Alarms

1. Follow your established procedure for an increasing combustible gas

condition.

2. Reset the alarm by pressing and releasing the POWER/MODE button

after the alarm condition has been cleared.

Responding to an Over Range Alarm

WARNING: An over range condition may indicate an extreme

combustible gas concentration. Confirm the gas
concentration with a different Model GP-01 or with
another gas detecting device.

1. Follow your established procedure for an increasing combustible gas

condition.

2. Reset the alarm by pressing and releasing the POWER/MODE button

Dead Battery Alarm

• Gas reading

replaced by FAIL.

• Battery icon

flashes.

• Double pulsing tone

(two pulses in quick
succession) occurs
once a second.

Sensor Failure

• Gas reading

replaced by FAIL.

• Double pulsing tone

(two pulses in quick
succession) once a
second.

System Failure

• Gas reading

replaced by FAIL.

SYS displays below

FAIL.

• Double pulsing tone

(two pulses in quick
succession) once a
second.

Table 4: Alarm Types and Indications

Alarm Type

LCD Indications

Other Indications