RKI Instruments 35-3001-01-03 User Manual

Page 9

Advertising
background image

35-3001-01-03 Methane Sample-Draw Detector • 5

door. The green Pilot LED is on when the sample-draw detector is receiving power from the
controller. The red Fail LED is on when the sample flow rate is below the low flow level.

Pressure Switch

The pressure switch is mounted to the back of the flowmeter circuit board. The pressure switch
monitors the flow rate of the incoming gas sample.

If the flow rate falls below the preset low flow level, the pressure switch causes the Fail LED to
turn on and interrupts the signal from the detector. The interrupted detector signal causes a fail
condition at the controller. The low flow level is factory-set at 0.6 SCFH (±0.1 SCFH). See
“Adjusting the Low Flow Setting” on page 15 to adjust this setting.

Pump Reset Switch

The pump reset switch is located to the left of the status LEDs. When a low flow condition occurs,
the pump will be shut off. To reset the low flow condition and start the pump again, press and hold
the pump reset switch for about 2 seconds, then release.

Hydrophobic Filter

The hydrophobic filter is located toward the bottom left of the main circuit board. The filter
prevents particulates and water in the incoming gas sample from damaging the flow and detection
systems. Replace the filter when it appears dirty, discolored, or clogged.

Pump

The pump is mounted to the right side of the main circuit board. The pump pulls the gas sample
into the sample-draw detector. The pump operates on 24 VAC, which is generated from the 24
VDC supplied by the controller.

IR Methane Sensor/IR Methane Sensor PCB

The methane sensor is an infrared type plug-in detector. Infrared combustible detectors are
generally used instead of the catalytic combustible detectors in applications where there may be
catalyst poisons such as silicone present or where oxygen is not present in the monitoring
environment.

The methane sensor is installed in the IR methane flow block and the IR methane flow block is
mounted to the middle of the main circuit board. A small circuit board, the IR methane sensor PCB,
with a 4-wire cable mates to the IR methane sensor and retains it in the flow block. The cable is
wired to the main PCB. The IR methane sensor PCB allows you to replace the sensor without
disconnecting the wiring.

Advertising