RKI Instruments 35-3001-02 User Manual

Page 18

Advertising
background image

14 • 35-3001-02 Oxygen Sample-Draw Detector

3.

If necessary, set the correct flow rate with the sensor flow control valve.

4.

If you cannot set the correct flow rate, check the sample line for obstructions or kinks.

5.

If the calibration/response difficulties continue, replace the sensor as described later in this
section.

6.

If the calibration/response difficulties continue, contact RKI Instruments, Inc. for further
instruction.

Replacing Components of the Sample-Draw Detector

This section includes procedures to replace the oxygen sensor, hydrophobic filter, and particle
filter.

Replacing the Oxygen Sensor

1.

Turn off the controller.

2.

Turn off power to the controller.

3.

Open the housing door of the sample-draw detector.

4.

Loosen the two screws on the oxygen terminal strip and remove the lugs and wires from the
terminal strip.

5.

Unscrew and remove the two screws that secure the sensor retaining plate, then lift the plate
and sensor out of the flow block.

There is a gasket at the bottom of the flow block. Be sure the gasket stays in place.

6.

Place the replacement sensor in the oxygen flow block, then position the retaining plate on the
two standoffs.

7.

Secure the retaining plate to the standoffs with the two screws you removed in step 4.

8.

Guide the wires and lugs of the replacement sensor to the oxygen terminal strip and insert each
lug into the appropriate terminal. See Figure 5 on page 10 for wiring connections. Tighten the
screws on the terminal strip.

9.

Close and latch the housing door.

10. Turn on power to the controller.

11. Turn on the controller.

CAUTION:

Allow the replacement sensor to warm up for 5 minutes before you continue.

12. Calibrate the replacement sensor as described in “Calibration” on page 16.

Replacing the Hydrophobic Filter

1.

Open the housing door of the sample-draw detector.

2.

Disconnect the filter from the rubber elbows on each end of the filter, then remove the filter
from the sample-draw detector.

3.

Install the new filter. Be sure the side of the filter marked “INLET” is connected to the elbow
that is connected to the inlet fitting.

4.

Verify that the flow rate is approximately 1.0 SCFH, then close the housing door.

Replacing the Particle Filter

1.

Disconnect the tubing routed to the sampling area, if installed, from the particle filter.

2.

Disconnect the particle filter from the inlet fitting’s tubing stub.

Advertising