Calibration – RKI Instruments 35-3001-02 User Manual

Page 20

Advertising
background image

16 • 35-3001-02 Oxygen Sample-Draw Detector

Calibration

This section describes how to calibrate the oxygen sensor in the sample-draw detector. It includes
procedures to prepare for calibration, set the fresh air reading, set the zero reading, and return to
normal operation.

NOTE: Calibrating the sample draw detector may cause alarms. Be sure to put the controller into

its calibration program or disable external alarms before continuing.

NOTE: This procedure describes calibration using a demand flow regulator.

Preparing for Calibration

NOTE: If you can verify a fresh air environment, it is not necessary to use a zero air calibrating

sample to set the fresh air reading at the controller.

1.

Follow the instructions in the controller’s operator’s manual for entering calibration mode.

2.

Screw the regulator into the zero air calibration cylinder.

Setting the Fresh Air Reading

1.

Connect the sample tubing from the demand flow regulator to the sample-draw detector’s inlet
fitting. This step is not necessary if you verified a fresh air environment earlier in this
procedure.

2.

Allow the sample draw detector to draw sample for one minute.

3.

Follow the directions in the controller’s operator’s manual for setting the fresh air reading.

4.

Disconnect the sample tubing from the inlet fitting.

5.

Unscrew the regulator from the zero air calibration cylinder.

Setting the Zero Reading

1.

Screw the demand flow regulator onto the 100% N

2

calibration cylinder.

2.

Connect the sample tubing from the regulator to the inlet line.

3.

Allow the sample-draw detector to draw the calibrating sample for 1 minute.

4.

Follow the directions in the controller’s operator’s manual for setting the zero (oxygen free)
reading.

5.

Disconnect the sample tubing from the sample-draw detector’s inlet fitting.

6.

Reconnect the incoming sample line to the inlet fitting.

7.

Unscrew the regulator from the calibration cylinder.

Returning to Normal Operation

1.

Wait 1 to 2 minutes to allow the oxygen reading at the controller to return to normal, then
return the controller to normal operation.

NOTE: If you do not allow the oxygen reading to return to normal, then unwanted alarms may

occur.

Advertising